by credit card, Paypal and bank transfer.
with BRT, DPD and FedEx couriers
via Whatsapp + 39.371.43.61.201
Resistant and compact structure, large high resolution LCD-TFT color display. Handle to facilitate movement. ECG, RESP, SpO2, NIBP, PR, 2-TEMP display. Indicated for use on adults and newborns. Arrhythmia analysis. Pacemaker detention, defibrillation and protection from electrosurgical interference. 96 hours of memory of all values, 500 NIBP measurements and 60 alarm events. USB port. Optional integrated printer.
Accessories supplied
- 1 5-way ECG cable with anti-defibrillation clip, IEC, 3.5m , reusable
- 30 ECG electrodes, adults, disposable
- 1 finger SPO2 sensor, adults, 2.5m , reusable
- 1 NIBP Cuff, Adult, 25cm-35cm, reusable
- 1 NIBP hose, 3m
- 1 Temperature Probe Skin, adult, 3m, reusable
- 1 Power Supply (EUR standard)
- 1 rechargeable lithium battery (14.8V, 2100mAh)
- 1 Earth cable
Languages available on request
- Italian
- English
- Spanish
- French
- Portuguese
- German
- Polish
- Chinese
- Turkish
Technical features
Display
Color TFT LCD: 10.1”
Resolution: 800x480
Tracks displayed: up to 8
Waveforms displayed: up to 11
Oscillation: 6.25, 12.5, 25, 50 mm/s
ECG
Leads: 5 leads and 3 selectable leads
Display sensitivity: Auto x 0.125, x 0.25, x 0.5, x 1, x 2, x 4
Scanning speed: 6.25, 12.5, 25, 50 mm/s
Heart Rate Range (HR):
Adults 15~300 bpm
Pediatric/Neonatal 15~350 bpm
Resolution: 1 bpm
Accuracy: ±1 bpm or 1%
Filter: Diagnosis mode 0.05 - 150 Hz / Monitoring mode 0.5 - 40 Hz / Surgical mode 1 - 20 Hz
Diagnosis mode 0.05 ~150 Hz
Monitoring mode 0.5~40 Hz
Surgical mode 1~20 Hz
Segment Tracking:
Measuring range -2.0 mV ~ 2.0 mV
Alarm range -2.0 mV ~ 2.0 mV
RESP
Method: trans-thoracic impedance
Operation: automatic/manual
RR measuring range:
Adults 0 ~150 rpm
Pediatric/Neonatal 0~150 bpm
Resolution: 1 rpm
Apnea alarm threshold:
10, 15, 20, 25, 30, 35, 40 s
Bandwidth: 0.2 - 2.5 Hz (-3 dB)
Scanning speed: 6.25, 12.5, 25, 50 mm/s
SpO2
Measurement and alarm range: 0~100%
Resolution: 1%
Precision:
±2% (70~100%, Adult, Pediatric)
±3% (70~100%, Neonatal)
Pulse measurement: 25~300 bpm
Resolution: 1 bpm
Update rate: 1 second
NIBP
Method: Automatic oscillometric
Operation: Manual/Automatic/Continuous
Measuring range in automatic mode:
1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 240, 480 minutes
Unit of measurement: mmHg/kPa
Type of measurement: Systolic, Diastolic, Average
Pressure range:
- Adults
Systolic 40~270 mmHg
Diastolic 10~215 mmHg
Average 20~235 mmHg
- Pediatric
Systolic 40~200 mmHg
Diastolic 10~150 mmHg
Average 20~165 mmHg
- Neonatal
Systolic 40~135 mmHg
Diastolic 10~100 mmHg
Average 20~110 mmHg
Resolution: 1mmHg
Accuracy: Max average error ± 5 mmHg
Max standard deviation: 8 mmHg
Pulse measurement range from NIBP: 40~240 bpm
Resolution: 1 bpm
Accuracy: 3 bpm or 3.5 %
Temperature (2 channels, 1 probe supplied)
Measurement/alarm range: 0~50°C
Resolution: 0.1°C
Accuracy: ±0.1°C (without probe)
Channel: 2
Printer (Optional)
Printing speed: 25, 50 mm/s
Diet
Power outlet: 100~240 V AC, 50/60 Hz
Rechargeable lithium battery: 14.8V
Dimensions L:32 x D:15 x H:26.5 cm